Medical News

GE Logiq S8 Image Interference

Views : 735
Update time : 2023-02-27 14:12:00
   Faulty Phenomenon:  The 2D image is normal, however there is obvious interference in color blood Image. 
 
 
Anaysis: Plug the probe in the different connectors.  The interface still exits. Check that the probe is normally used on other machines, so it is not the probe problem. It is judged that the BF192 board is faulty.
 
Troubleshooting:After repair the BF192 board, the image is back to normal.
 
 
Related News
Read More >>
Hitachi Aloka HI VISION PREIRUS: Image Interference Hitachi Aloka HI VISION PREIRUS: Image Interference
Jan .31.2024
Image Interference Troubleshooting for Hitachi Aloka HI VISION Preirus
GE VIVID I system maintainance GE VIVID I system maintainance
Jan .30.2024
Ultrasound system,Hospital equipment,Medical diagnosis implement,Imagine parts,Medical spare parts,Ultrasonic Sensor,Hitachi Aloka,GE,Siemens,Toshiba,Philips
About Us
As one of the leading ultrasound service companies in China, Rongtao Medical provides healthcare solutions, repair service, parts and accessories to hospitals, clinics and distributors world-widely.
Connect with us

Copyright 2014-2023 Guangzhou Rongtao Medical Tech LTD All Rights Reserved.