Medical News

Repair case: SIEMENS Acuson Antares

Views : 528
Update time : 2021-03-19 12:04:06
                                                                               

The fault phenomenon
After the Antares machine starts, the display stays at the SIEMENS interface, and the machine restarts automatically repeatedly.

Failure analysis
After testing the other components without any problems, the DVD driver was replaced, the system software was successfully reinstalled, the machine was restarted after the system disc automatically exited, and the firmware was loaded into the ultrasonic two-dimensional interface and could be used normally.
 
 
Summary
Replacement of hard disk, DVD drive, reinstallation of system software, normal start-up of the machine, troubleshooting.




Related News
Read More >>
Hitachi Aloka HI VISION PREIRUS: Image Interference Hitachi Aloka HI VISION PREIRUS: Image Interference
Jan .31.2024
Image Interference Troubleshooting for Hitachi Aloka HI VISION Preirus
GE VIVID I system maintainance GE VIVID I system maintainance
Jan .30.2024
Ultrasound system,Hospital equipment,Medical diagnosis implement,Imagine parts,Medical spare parts,Ultrasonic Sensor,Hitachi Aloka,GE,Siemens,Toshiba,Philips
About Us
As one of the leading ultrasound service companies in China, Rongtao Medical provides healthcare solutions, repair service, parts and accessories to hospitals, clinics and distributors world-widely.
Connect with us

Copyright 2014-2023 Guangzhou Rongtao Medical Tech LTD All Rights Reserved.