Medical News

The GE LOGIQ S8 fails to start

Views : 406
Update time : 2022-01-04 16:16:00

Failure phenomenon:

When GE LOGIQ S8 is started, the startup light is on, but no signal is displayed, so it cannot be started normally.

 

Failure analysis:

After the machine presses the on key, the display shows no signal. Consider the power supply and motherboard, and troubleshoot according to the carried accessories. After replacing the power supply, the machine can start. However, after each startup, the system reports an error, which still needs further investigation. After continuing to replace the motherboard, the system still reports an error when starting up. Unplug the BF board again and eliminate the fault.

 

Repair results:

After replacing the power supply and removing and inserting the BF board, the fault is removed and the machine is in normal use.

 

 

Related News
Read More >>
Hitachi Aloka HI VISION PREIRUS: Image Interference Hitachi Aloka HI VISION PREIRUS: Image Interference
Jan .31.2024
Image Interference Troubleshooting for Hitachi Aloka HI VISION Preirus
GE VIVID I system maintainance GE VIVID I system maintainance
Jan .30.2024
Ultrasound system,Hospital equipment,Medical diagnosis implement,Imagine parts,Medical spare parts,Ultrasonic Sensor,Hitachi Aloka,GE,Siemens,Toshiba,Philips
About Us
As one of the leading ultrasound service companies in China, Rongtao Medical provides healthcare solutions, repair service, parts and accessories to hospitals, clinics and distributors world-widely.
Connect with us

Copyright 2014-2023 Guangzhou Rongtao Medical Tech LTD All Rights Reserved.