GE VOLUSON E10:Image is whitish
 

Failure phenomenon
The scan area image of GE VOLUSON E10 color ultrasound system is whitish, and restarting is invalid.


Troubleshooting
Voluson E10 image problems are common front-end board problems.
Diagnose according to the image conditions, replace the RFM board, and troubleshoot after reinstalling the system.




Tips:
Poor contact between the various components of the ultrasound equipment is a common cause of failure of the ultrasound system. Dust pollution, contact oxidation, frequent insertion of probes, bumps during use, etc. may all lead to poor contact. Regular dust removal, regular maintenance and preventive maintenance if possible, avoiding human damage as much as possible is an effective way to reduce the failure of the ultrasound system.