ultrasound transducer probe, ultrasound repair service, medical repair

ultrasonic board repair
Categories
Support
+
Philips IE33 ultrasonic diagnostic instrument common fault troubleshooting(6)


Fault phenomenon
The probe is not fully connected to the device, resulting in loss of crystal oscillator signal or other image quality problems.

Troubleshooting
The inner hexagon screw used for internal fixation of probe interface has slipped and the probe cannot be fully connected with the probe interface module on the machine.Use inch 3/32 socket head tool with set screws.
Message
If you have any suggestions or question for us.Please contact us.
*
*
*
*
*
About Us
As one of the leading ultrasound service companies in China, Rongtao Medical provides healthcare solutions, repair service, parts and accessories to hospitals, clinics and distributors world-widely.
Connect with us

Copyright 2014-2023 Guangzhou Rongtao Medical Tech LTD All Rights Reserved.