Toshiba SSA-580A Maintenance

Fault phenomenon
The machine starts normally, but there is no echo when it enters the ultrasonic working interface.
 
Fault analysis
Ultrasound image echo-free is a typical fault that all major brands of ultrasound machines will encounter. Usually, the reason for the failure is that the front-end plate of the power supply of the machine has no high-voltage output or is responsible for the processing function of ultrasound imaging. It is preliminary judged that there may be faults in the A43 FETX board and the A60 ETXR2 board, among which the A43 FETX board is the transmitting and receiving board and the A60 ETXR2 board is responsible for the tube of the high-voltage board. High voltage output of the probe.
 
Fault summary
After replacing the A60 ETXR2 board, the machine starts up normally. After entering the ultrasonic working interface, the image echo area is restored and the machine runs normally.


Rongtao Medical's mission is to help people realize the efficient circulation utilization of medical resource!
We equipped with:
                              1.Professional&Passionate pre-sale service team
                              2.10 years experience repair technician team
                              3.Reliable&Responsible quality control team
                              4.Experienced&Specialize after-sale technical support team

Our vision : TO be the top preferred third-party service company worldwide!
                            
Message
If you have any suggestions or question for us.Please contact us.
*
*
*
*
*
About Us
As one of the leading ultrasound service companies in China, Rongtao Medical provides healthcare solutions, repair service, parts and accessories to hospitals, clinics and distributors world-widely.
Connect with us

Copyright 2014-2023 Guangzhou Rongtao Medical Tech LTD All Rights Reserved.