ultrasound transducer probe, ultrasound repair service, medical repair

ultrasonic board repair
Categories
Support
+
Philips IE33 ultrasonic diagnostic instrument common fault troubleshooting(2)
 
Fault phenomenon
System error"602"

Troubleshooting
Most of the 602 errors are due to memory: loose connections between memory and slots, or poor contact due to oxidation or dirt on the surface of memory gold fingers.First eliminate 602 errors due to incorrect BIOS memory hole setup (use FII Instruction to verify the hard-ware /Software memory hole setup for the current Software version).After cleaning and refixing the memory, the equipment started normally, troubleshooting.
Message
If you have any suggestions or question for us.Please contact us.
*
*
*
*
*
About Us
As one of the leading ultrasound service companies in China, Rongtao Medical provides healthcare solutions, repair service, parts and accessories to hospitals, clinics and distributors world-widely.
Connect with us

Copyright 2014-2023 Guangzhou Rongtao Medical Tech LTD All Rights Reserved.