                                                                                  GE VOLUSON E8: Blue screen
 

Fault phenomenon

The GE VOLUSON E8 color ultrasound system has a blue screen and reports that the system file is missing. It is resolved after reinstalling the system, but this failure recurs after a week.

 

Troubleshooting

According to the description, reinstalling the system has solved the fault, indicating that the computer system software crashed, but the same fault occurred a week after it returned to normal. It is suspected that the hard disk may have bad sectors, causing the system file to be lost during startup and causing the fault.

Open the panel, remove the hard disk, find a hard disk with the same capacity, use Ghost to clone a hard disk with the operating system and ultrasound application software using the method of hard disk engraving, complete the hard disk replacement, and troubleshoot.