Philips IE33 ultrasonic diagnostic instrument common fault troubleshooting(3)
Fault phenomenon
The device failed to start, stopped at the "initializing hardware" screen, or error 606 was reported during startup.
Troubleshooting
The motherboard battery is dead and the BIOS Settings will automatically jump to the Default Settings, which will cause an exception to the HOST or SIP boot.Mainboard battery life is about 5 years, use a multimeter to measure the battery voltage, less than +2.85Vdc, replace the mainboard battery (HOST/SIP mainboard battery model is CR2032, UMB/EMB mainboard battery model is CR2450), and change the BIOS to Philips Settings, troubleshooting.
Message
If you have any suggestions or question for us.Please contact us.