Key service context for quote intake and technical review
Published in the Rongtao Medical Blog under Repair Guide.
Relevant system context: GE LOGIQ 500.
Technical focus: Probe / Transducer.
Part reference: 2349040-23.
Use this page to compare product, repair and troubleshooting context before opening a quote request.
BrandGE
Model / systemLOGIQ 500 · LOGIQ 500PRO · LOGIQ 200E
Part typeProbe / Transducer
Part number2349040-23
Content typeRepair Guide
Technical notes
Technical body content
No image when scanning far field
Check power supply, r einsert the circuit board . Suggest to do metal earthing.
LOGIQ 200E
Curved lines interpolated into sector-shaped image, from near field to far field
L ithium battery - short life
Replace lithium battery , and upgrade the system to 2349040-23 KB824146
LOGIQ 400MD
cardiac probe shows clutter filte , interference in far field
Service context
Rongtao uses this technical record for GE LOGIQ 500 Probe / Transducer as supporting context for quote intake, photo comparison, board identification and repair-route selection. For live availability, send the machine model, board label, fault symptom and clear photos of both sides of the part.