Fault phenomenon
The S2000 machine shuts down automatically about half an hour after normal use, and there is no error warning during shutdown.
F ault analysis
The automatic shutdown of color Doppler machine in use is usually caused by the failure of power supply or the poor heat dissipation effect of the machine. Due to the normal start-up process of the machine, first consider whether the excessive accumulation of dust inside the machine leads to the decline of the heat dissipation capacity of the machine. After opening the case, a large amount of dust accumulated in the case is found. After dust removal and cleaning, the machine is restarted for testing. The machine will shut down automatically again after one hour of use. Therefore, it is judged that the failure should be related to the heat dissipation effect of the machine and the power supply part appears. Obstacles are less likely. Because this fault is similar to the general computer heat dissipation fault, the general computer will shut down or shut down automatically when the CPU temperature of the main board is too high, and the CPU temperature is too high, which is caused by the failure of the cooling fan or the power supply circuit of the main board itself. The RM board in the S2000 machine is similar to the computer motherboard. It is tested by replacing the board. It feels hot when touching the RM board, which further confirms the possibility of poor heat dissipation effect of the RM board.
Troubleshooting
After replacing the RM board, the automatic shutdown failures did not occur again in the afternoon of boot-up test. Troubleshooting. Rongtao Medical's mission is to help people realize the efficient circulation utilization of medical resource!
We equipped with
1.Professional&Passionate pre-sale service team 2.10 years experience repair technician team 3.Reliable&Responsible quality control team 4.Experienced&Specialize after-sale technical support team Our vision : TO be the top preferred third-party service company worldwide!
Service context
Rongtao uses this technical record for GE VIVID ULTRASOUND Probe / Transducer as supporting context for quote intake, photo comparison, board identification and repair-route selection. For live availability, send the machine model, board label, fault symptom and clear photos of both sides of the part.