Fault phenomenon
When the GE LOGIQ E9 color ultrasound pressed the unlock button, the operating keyboard was still locked and could not be moved.
Failure analysis
After the lock repair and calibration is done in the system service program, the fault remains the same; Remove the motor locking mechanism in the fault leg, press the test button after starting up, the motor rotates normally, but the locking slider cannot move at the end, so it is judged to be the locking slider fault.
Troubleshooting
When the device is powered on, press the unlock button to ensure motor rotation. At the same time, release the locking slider with a special tool and reinstall the device.Enter the system service, do lock repair calibration, restart the system after the normal.