Fault phenomenon
The machine starts normally, but there is no echo when it enters the ultrasonic working interface.
F ault analysis
Ultrasound image echo-free is a typical fault that all major brands of ultrasound machines will encounter. Usually, the reason for the failure is that the front-end plate of the power supply of the machine has no high-voltage output or is responsible for the processing function of ultrasound imaging. It is preliminary judged that there may be faults in the A43 FETX board and the A60 ETXR2 board, among which the A43 FETX board is the transmitting and receiving board and the A60 ETXR2 board is responsible for the tube of the high-voltage board. High voltage output of the probe.
Fault summary
After replacing the A60 ETXR2 board, the machine starts up normally. After entering the ultrasonic working interface, the image echo area is restored and the machine runs normally. Rongtao Medical's mission is to help people realize the efficient circulation utilization of medical resource!
We equipped with
1.Professional&Passionate pre-sale service team 2.10 years experience repair technician team 3.Reliable&Responsible quality control team 4.Experienced&Specialize after-sale technical support team Our vision : TO be the top preferred third-party service company worldwide!
Service context
Rongtao uses this technical record for GE VIVID ULTRASOUND Probe / Transducer as supporting context for quote intake, photo comparison, board identification and repair-route selection. For live availability, send the machine model, board label, fault symptom and clear photos of both sides of the part.