GE Voluson E10 · RFM — Radio-Frequency Module
Case 001 · TAT 6 days · 5 photos on file
Fault
B-mode image anomaly across all probes
Fix
Beamformer module fault isolated to RFM. Replaced RFM; image returned to spec on warm + cold boot.
Service narrative

Persistent imaging anomaly across multiple probes. Bench-isolated to the RFM beamformer module. Replaced the failing front-end ICs, recalibrated the full channel map, and reverified each probe class on a live Voluson E10 chassis for 48 hours before sign-off.

Case record
Case number001
BrandGE
ModelVoluson E10
BoardRFM — Radio-Frequency Module
TAT6 days
QA protocol48 h real-device
Warranty60 days · return shipping covered