B2B ultrasound repair & partsGlobal shippingEmail quote intakeNeed help? info@rongtaomedical.com
Rongtao Medical
RONGTAO MEDICAL
ULTRASOUND REPAIR · MODEL PAGES · LIVE INVENTORY
Contact Us
GE Voluson E10 · RFM423 — Radio-Frequency Module
Case 010 · TAT 7 days · 3 photos on file
Fault
All probes return no echo on startup
Fix
RFM423 channel rework; element-by-element calibration. Echo restored across linear, convex and 4D probes.
Service narrative

Persistent imaging anomaly across multiple probes. Bench-isolated to the RFM beamformer module. Replaced the failing front-end ICs, recalibrated the full channel map, and reverified each probe class on a live Voluson E10 chassis for 48 hours before sign-off.

Case record
Case number010
BrandGE
ModelVoluson E10
BoardRFM423 — Radio-Frequency Module
TAT7 days
QA protocol48 h real-device
Warranty60 days · return shipping covered
Service photo log
3 bench photos on file
010-01Service photo
010-02Service photo
010-03Service photo
‹ Case 009Voluson E10⊞ All casesCase 011Voluson E10