Fault
System will not power on
Fix
PSU rebuild — primary rail caps + control IC replaced. 48 h real-device burn-in cleared.
Service narrative

Classic no-power fault. Bulk caps had failed; rebuilt the supply, replaced all suspect electrolytics, verified every rail under load and burned in for 48 hours on a live chassis.

Case record
Case number013
BrandGE
ModelVoluson E8
BoardMain power supply
TAT5 days
QA protocol48 h real-device