ge logiq s8 2d image and blood flow abnorma
Repair Cases · Repair case · Reviewed May 18, 2026 · A00149A1
Repair caseText-focused technical page

This page is focused on the technical note, model context and service information rather than product photography.

Page summary
Key service context for quote intake and technical review
  • Published in the Rongtao Medical Blog under Repair Case.
  • Relevant system context: GE LOGIQ S8.
  • Technical focus: Probe / Transducer.
  • Part reference: A00149A1.
  • Use this page to compare product, repair and troubleshooting context before opening a quote request.
BrandGE
Model / systemLOGIQ S8 · VIVID ULTRASOUND · VIVID FIRST
Part typeProbe / Transducer
Part numberA00149A1
Content typeRepair Case
Technical notes
Technical body content

First, the failure phenomenon

2D image and blood flow of the machine is abnormal , the blood flow knob is not working good .

Second, the failure analysis

According to the above failure phenomenon, it can be judged that the image channel of the machine is abnormal. Therefore, it is suspected that there is a problem with the BF channe l board. It was also found that the knob of blood flow section was not working goog , so the control panel was disassembled and the circuit board of the blood flow button needs to be repaired .

Third, testing and repair

After replace the BF channel board and repair the circuit board of the blood flow button , everything back to normal.

Service context

Rongtao uses this technical record for GE LOGIQ S8 Probe / Transducer as supporting context for quote intake, photo comparison, board identification and repair-route selection. For live availability, send the machine model, board label, fault symptom and clear photos of both sides of the part.