GE LOGIQ S8: BF and control panel failure
Repair Cases · Repair case · Reviewed May 18, 2026 · A00144A1
Repair caseText-focused technical page

This page is focused on the technical note, model context and service information rather than product photography.

Page summary
Key service context for quote intake and technical review
  • Published in the Rongtao Medical Blog under Repair Case.
  • Relevant system context: GE LOGIQ S8.
  • Technical focus: Control Panel.
  • Part reference: A00144A1.
  • Use this page to compare product, repair and troubleshooting context before opening a quote request.
BrandGE
Model / systemLOGIQ S8
Part typeControl Panel
Part numberA00144A1
Content typeRepair Case
Technical notes
Technical body content

Failure phenomenon

The two-dimensional image of the machine is abnormal, and the blood flow is abnormal; the blood flow knob is not sensitive.

Troubleshooting

According to the above failure phenomenon, it can be judged that the image channel of the machine is abnormal. Therefore, it is suspected that there is a problem with the BF board of the machine, so replace the BF board. After replacing the BF, remove the fault.

Disassembly steps

1. First, remove the BF board from the back of the machine. 2. Next, replace the BF board with a new one. 3. Then, turn on the machine again, and the image displays normally. 4. It was found that the blood flow knob was not sensitive during the malfunction of the machine, so the control panel of the machine was disassembled and the circuit board where the blood flow button was replaced was performed. 5. Result: After replacing the circuit board where the blood flow button is located, the knob returns to normal.

Repair result

After replacing the BF board and the circuit board where the blood flow button is located, the machine malfunction is resolved.

Service context

Rongtao uses this technical record for GE LOGIQ S8 Control Panel as supporting context for quote intake, photo comparison, board identification and repair-route selection. For live availability, send the machine model, board label, fault symptom and clear photos of both sides of the part.

Tags and related links
Current internal destinations only
Related pages
Same category, brand, model or tag context
Repair case
Repair Cases · Repair case

GE VOLUSON E10:Image is whitish

GE, LOGIQ P5, Beamformer / RF Board, PN A00142A1 repair case with part identification, model context and Rongtao quote or repair support.

Reviewed May 18, 2026A00142A1
Repair case
Repair Cases · Repair case

the ge logiq s8 fails to start

GE, LOGIQ S8, Probe / Transducer, PN A00148A1 repair case with part identification, model context and Rongtao quote or repair support.

Reviewed May 18, 2026A00148A1