GE VIVID I system maintainance
Repair Guides · Technical article · Updated Jan 30, 2024 · I00004I1
Technical articleText-focused technical page

This page is focused on the technical note, model context and service information rather than product photography.

Page summary
Key service context for quote intake and technical review
  • Published in the Rongtao Medical Blog under Article.
  • Relevant system context: GE VIVID I.
  • Part reference: I00004I1.
  • Use this page to compare product, repair and troubleshooting context before opening a quote request.
BrandGE
Model / systemVIVID I
Part numberI00004I1
Content typeArticle
Technical notes
Technical body content

First, the failure phenomenon

The machine starts normally, and after receiving the probe, it prompts an error and the probe is not recognized.

Second, the failure analysis

The ultrasonic probe does not recognize the power supply that normally considers the probe interface board or supplies power to the probe interface board. The machine configuration diagram shows that there is no high-voltage switch on the interface board of the model, that is, it does not affect the high-voltage output of the DC power supply, so it can pass the detection. The DC output voltage directly determines whether there is a fault in the power supply. After the detection, the machine is found to have no high-voltage output, and it is determined that the DC power supply is faulty.

Third, testing and repair

Replace the DC-DC power supply, the probe can be recognized normally after the machine is started, and each function resumes normal operation. In genneral, we are really experted in ultrasound repair.

Tags and related links
Current internal destinations only
Related pages
Same category, brand, model or tag context
GE Logiq S8 Image Interference
Repair Guides · Technical article

GE Logiq S8 Image Interference

GE, LOGIQ S8, PN I00624I1 technical article with photos, part identification, model context and Rongtao quote or repair support.

Updated Feb 27, 2023I00624I12 images
Technical article
Repair Guides · Technical article

GE Logiq P5

GE, LOGIQ P5, PN I00196I1 technical article with part identification, model context and Rongtao quote or repair support.

Updated Dec 1, 2020I00196I1