Philips IE33 ultrasonic diagnostic instrument common fault troubleshooting(2)
Troubleshooting
Most of the 602 errors are due to memory: loose connections between memory and slots, or poor contact due to oxidation or dirt on the surface of memory gold fingers.First eliminate 602 errors due to incorrect BIOS memory hole setup (use FII Instruction to verify the hard-ware /Software memory hole setup for the current Software version).After cleaning and refixing the memory, the equipment started normally, troubleshooting.
Service context
Rongtao uses this technical record for Philips VIVID ULTRASOUND Probe / Transducer as supporting context for quote intake, photo comparison, board identification and repair-route selection. For live availability, send the machine model, board label, fault symptom and clear photos of both sides of the part.