Philips IE33 ultrasonic diagnostic instrument common fault troubleshooting(5)
Fault phenomenon
Images imported from external media into i-33 cannot be re-exported in any format.
Troubleshooting
The problem is that the system does not allow Imported checks to be exported.Open the import check then Double-click the first image to go to full screen mode. Press the Acquire key on I plus 33, which will Append the check. After that, open each imported image to the full screen, press the Acquire key on positive 33, so that the image can be copied to the new Append check below.Close the Re - view;Open the Review interface, select appended study, and then select Send to Media to export.Open check can also be output in Jpeg/AVI format, problem solved.
Service context
Rongtao uses this technical record for GE VIVID ULTRASOUND Probe / Transducer as supporting context for quote intake, photo comparison, board identification and repair-route selection. For live availability, send the machine model, board label, fault symptom and clear photos of both sides of the part.