B2B ultrasound repair & partsGlobal shippingEmail quote intakeNeed help? info@rongtaomedical.com
Rongtao Medical
RONGTAO MEDICAL
ULTRASOUND REPAIR · MODEL PAGES · LIVE INVENTORY
Contact Us
philips ie33 ultrasonic diagnostic instrument common fault troubleshooting5
Brand Knowledge · Repair guide · Reviewed May 18, 2026
Repair guideText-focused technical page

This page is focused on the technical note, model context and service information rather than product photography.

Page summary
Key service context for quote intake and technical review
  • Published in the Rongtao Medical Blog under Repair Guide.
  • Relevant system context: GE VIVID ULTRASOUND.
  • Technical focus: Probe / Transducer.
  • Use this page to compare product, repair and troubleshooting context before opening a quote request.
BrandGE
Model / systemVIVID ULTRASOUND · VIVID PHILIPS · EPIQ
Part typeProbe / Transducer
Content typeRepair Guide
Technical notes
Technical body content

Technical notes

Philips IE33 ultrasonic diagnostic instrument common fault troubleshooting(5)

Fault phenomenon

Images imported from external media into i-33 cannot be re-exported in any format.

Troubleshooting

The problem is that the system does not allow Imported checks to be exported.Open the import check then Double-click the first image to go to full screen mode. Press the Acquire key on I plus 33, which will Append the check. After that, open each imported image to the full screen, press the Acquire key on positive 33, so that the image can be copied to the new Append check below.Close the Re - view;Open the Review interface, select appended study, and then select Send to Media to export.Open check can also be output in Jpeg/AVI format, problem solved.

Service context

Rongtao uses this technical record for GE VIVID ULTRASOUND Probe / Transducer as supporting context for quote intake, photo comparison, board identification and repair-route selection. For live availability, send the machine model, board label, fault symptom and clear photos of both sides of the part.

Tags and related links
Current internal destinations only
Related pages
Same category, brand, model or tag context
replace philips hd15 aim board
Brand Knowledge · Repair guide

replace philips hd15 aim board

GE, VIVID ULTRASOUND, Probe / Transducer, PN 453561197306 repair guide with photos, part identification, model context and Rongtao quote or repair support.

Reviewed May 18, 20264535611973061 images