B2B ultrasound repair & partsGlobal shippingEmail quote intakeNeed help? info@rongtaomedical.com
Rongtao Medical
RONGTAO MEDICAL
ULTRASOUND REPAIR · MODEL PAGES · LIVE INVENTORY
Contact Us
ge logiq3 fault maintenance case 2
Repair Cases · Repair guide · Reviewed May 18, 2026
Repair guideText-focused technical page

This page is focused on the technical note, model context and service information rather than product photography.

Page summary
Key service context for quote intake and technical review
  • Published in the Rongtao Medical Blog under Repair Guide.
  • Relevant system context: GE LOGIQ3.
  • Technical focus: Probe / Transducer.
  • Use this page to compare product, repair and troubleshooting context before opening a quote request.
BrandGE
Model / systemLOGIQ3 · VIVID ULTRASOUND · VIVID GE
Part typeProbe / Transducer
Content typeRepair Guide
Technical notes
Technical body content

The fault phenomenon

GE LOGIQ3 machine boot display start working but the screen still didn ’ t works and show nothing,control panel indicator light on.

Failure analysis

The machine is not connected to the workstation, so it cannot be judged by the output image observation of the workstation. When the external display is started, the same phenomenon of black screen is found, and the fault of the display itself is eliminated.Since the control panel indicator keeps working during the startup process, the possibility of power failure is small, and the PC part of the machine is suspected to be faulty. PC chassis, after open the machine casing and the motherboard after connecting the PC power supplies and computer hard disk, at the same time connect the computer monitor output image, separate start the motherboard and found no display, then make sure it ’ s the failure of the mainboard, after the replacement motherboard and start LOGIQ3 machine, the machine is still show the same failure phenomenon, due to the change on the system board still use the original machine CPU and memory chips, when use a piece of embedded CPU mainboard after start the machine, the machine normal boot, display normal output ultrasonic work interface, thus concludes it ’ s the motherboard CPU machine fault.However, when starting the test the next day, the machine appeared blue screen, and then started the machine again appeared black screen failure, repeatedly starting test found that the machine sometimes can start normally, sometimes appeared black screen, failure appeared irregularly.As the machine appears black screen again, it occurs after the machine starts up with a blue screen, and the blue screen phenomenon is generally caused by the failure of the machine peripherals or some PC hardware or the conflict between the hardware and software of the machine, which is commonly a memory chip fault.

Troubleshooting

Purchase the same type of memory chips for testing, when the machine starts , no longer exist the black screen fault, then fault remov ed . To achieve Rongtao Medical 's goal of being the top prefered third-party service company worldwide.

We equipped with

★Professional & passionate pre-sale service team ★10 years experienced repair technician team ★Reliable & responsible quality control team ★Experienced & specialize after-sale technical support team Besides technical service, we have wide range of ultrasound parts/accessories, ultrasound probes, patient monitor parts, patient monitor models readily available for sale, not only to serve our customer with the best quality repairs, but the cost-effective sales. If you are looking for any related repair service or parts, please don’t hesitate to contact us, we are always here to help!

Service context

Rongtao uses this technical record for GE LOGIQ3 Probe / Transducer as supporting context for quote intake, photo comparison, board identification and repair-route selection. For live availability, send the machine model, board label, fault symptom and clear photos of both sides of the part.