Failure phenomenon

Philips CX50 report error 0022, can not enter the system.

Failure analysis

When the machine starts up, error 0022 is reported, which may be caused by the failure of the power board or channel board. First enter the background self-check, and the self-check result is the problem of the Pwr Sup power board, so the next troubleshooting of the power board.

Repair results

Replace the power supply, all the functions of the machine are used normally, and the fault is eliminated.