Philips EPIQ 7: Shows Error "511"

Repair Guides · Technical article · Updated Jul 8, 2022 · I00152I1

Page summary

Key service context for quote intake and technical review
  • Published in the Rongtao Medical Blog under Article.
  • Relevant system context: Philips EPIQ 7.
  • Part reference: I00152I1.
  • Use this page to compare product, repair and troubleshooting context before opening a quote request.
BrandPhilips
Model / systemEPIQ 7 · epiq
Part numberI00152I1
Content typeArticle

Technical notes

Technical body content

Fault Phenomenon

PHILIPS EPIQ 7 ultrasound power was suddenly cut off during use, and then switched on again after power was switched on. After two sudden power cuts, it started up again with abnormal start fault. The display stayed at the interface of "PHILIPS" for a long time, and finally entered the interface with an error of "511", the device couldn't be normally used.

Troubleshooting and maintenance

According to the description of the operator, to plug EPIQ7 according to boot key boot after Lord of ultrasonic power, found the screen for a long time stay is the BIOS testing interface (see below), restart enter BIOS interface, check the BIOS Settings appear problem, not on the basis of the BIOS test only for EPIQ7 front in self-inspection on startup, thus eliminate front-end equipment has a problem, should appear in front of the partial suspect malfunction. Pull out the ACQ and Power REG PCB in the back-end PC,remove all the front-end load, leaving the PC part only. The fault was still found, and it was determined that the fault should be in the PC part. Enter the startup interface and report the error "511". Check the error log and find the message "Plantform Monitor failed to start the Digital Media Manger process", that is, some parts are considered to have failed to start, so it is necessary to find the parts that have failed to start to determine whether they are the cause of the fault. Remove all circuit boards and cables connected to the PC motherboard, leaving only the hard disk, memory and graphics card. Restart EPIQ7 ultrasonic and it will be found that the BIOS detection interface will soon complete the detection, boot normally without any error, and the fault is eliminated. It is judged that the damaged part is inserted into the part just removed one at a time and then the device is restarted. It is found that the fault returns after the DVD cable is inserted into the motherboard and the fault still exists after the REPLACEMENT of the DVD cable. It is suspected that the DVD damage caused the EPIQ7 fault.

Tags and related links

Current internal destinations only

Related pages

Same category, brand, model or tag context
Repair Guides · Technical article

Philips EPIQ 7C:Time is too long to start up

Philips, EPIQ 7C, PN I00198I1 technical article with part identification, model context and Rongtao quote or repair support.

Updated Jan 31, 2024I00198I1
Repair Guides · Technical article

Fault case:Philips HD15 Ultrasound System

Philips, HD15, PN I00281I1 technical article with part identification, model context and Rongtao quote or repair support.

Updated Mar 19, 2021I00281I1
Repair Guides · Technical article

Philips CX50 report error 0022, can not enter the system

Philips, CX50, PN 0022-CAN-NOT-ENTER-THE-SYSTEM-I00473I1 technical article with part identification, model context and Rongtao quote or repair support.

Updated Feb 27, 20230022-CAN-NOT-ENTER-THE-SYSTEM-I00473I1
Repair Guides · Technical article

Philips IU22: Abdominal probe PW mode error

Philips, iU22, Probe / Transducer, PN I00236I1 technical article with part identification, model context and Rongtao quote or repair support.

Updated Feb 28, 2022I00236I1