GE VOLUSON E8: Blue screen

Repair Cases · Repair case · Reviewed May 18, 2026 · A00143A1

Page summary

Key service context for quote intake and technical review
  • Published in the Rongtao Medical Blog under Repair Case.
  • Relevant system context: GE LOGIQ E.
  • Part reference: A00143A1.
  • Use this page to compare product, repair and troubleshooting context before opening a quote request.
BrandGE
Model / systemLOGIQ E · VOLUSON E8 · VOLUSON E10
Part numberA00143A1
Applicationsoftware using the method of hard disk engraving, complete the hard disk replacement, and
Content typeRepair Case

Technical notes

Technical body content

Fault phenomenon

The GE VOLUSON E8 color ultrasound system has a blue screen and reports that the system file is missing. It is resolved after reinstalling the system, but this failure recurs after a week.

Troubleshooting

According to the description, reinstalling the system has solved the fault, indicating that the computer system software crashed, but the same fault occurred a week after it returned to normal. It is suspected that the hard disk may have bad sectors, causing the system file to be lost during startup and causing the fault. Open the panel, remove the hard disk, find a hard disk with the same capacity, use Ghost to clone a hard disk with the operating system and ultrasound application software using the method of hard disk engraving, complete the hard disk replacement, and troubleshoot.

Service context

Rongtao uses this technical record for GE LOGIQ E as supporting context for quote intake, photo comparison, board identification and repair-route selection. For live availability, send the machine model, board label, fault symptom and clear photos of both sides of the part.

Tags and related links

Current internal destinations only

Related pages

Same category, brand, model or tag context
Repair CasesRepair case

GE VOLUSON E10:Image is whitish

GE, LOGIQ P5, Beamformer / RF Board, PN A00142A1 repair case with part identification, model context and Rongtao quote or repair support.

Reviewed May 18, 2026A00142A1