B2B ultrasound repair & partsGlobal shippingEmail quote intakeNeed help? info@rongtaomedical.com
Rongtao Medical
RONGTAO MEDICAL
ULTRASOUND REPAIR · MODEL PAGES · LIVE INVENTORY
Contact Us
philips epiq 7c v6 2 probe artifact
Repair Cases · Repair case · Reviewed May 18, 2026 · A00134A1
Repair caseText-focused technical page

This page is focused on the technical note, model context and service information rather than product photography.

Page summary
Key service context for quote intake and technical review
  • Published in the Rongtao Medical Blog under Repair Case.
  • Relevant system context: GE VIVID ULTRASOUND.
  • Technical focus: Probe / Transducer.
  • Part reference: A00134A1.
  • Use this page to compare product, repair and troubleshooting context before opening a quote request.
BrandGE
Model / systemVIVID ULTRASOUND · VIVID PHILIPS · epiq 7c
Part typeProbe / Transducer
Part numberA00134A1
Content typeRepair Case
Technical notes
Technical body content

Failure phenomenon

The 2D image of the Philips EPIQ 7C ultrasonic three-dimensional volume probe V6-2 has an artifact on the right side of the center, and the artifact is more obvious after the image is colored.

Troubleshooting

First swap the slots between V6-2 and C5-1, and observe that the image of C5-1 is normal, and the artifacts in the middle right of the image of V6-2 still exist. It is determined that it is not the problem of the slot interface. Then there may be bubbles in the probe V6-2, select V6-2 for the probe, turn on the 4D mode, let the motor in the probe rotate for a period of time, then switch back to the 2D mode, and hang the probe upside down for a period of time. If there are bubbles, the above two steps are mainly It is to eliminate the bubbles in front of the array element. But image artifacts still exist after the above two steps.Moreover, the artifacts are elongated after painting. In the end, it may be that an array element in the probe is broken. Freeze the machine. Unplug all the probes and U disk to enter the maintenance service interface and perform probe test under Hardware Utilities. Copy out the test log (log) and analyze it. It is found that there are four elements in the probe damaged. In engineering practice, the energy of a single element is small when it is launched. Generally, multiple elements are combined to transmit, which plays a role in controlling the signal. Next, each array element is grouped and combined and excited in a certain way to form different emission beam scanning. If the array element is damaged, part of the image cannot be scanned. The device image returned to normal after the probe was replaced, and there was no failure after a period of trial.

Service context

Rongtao uses this technical record for GE VIVID ULTRASOUND Probe / Transducer as supporting context for quote intake, photo comparison, board identification and repair-route selection. For live availability, send the machine model, board label, fault symptom and clear photos of both sides of the part.