philips iu22 cannot start after ultrasonic power off
Repair Cases · Repair case · Reviewed May 18, 2026 · A00116A1
Repair caseText-focused technical page

This page is focused on the technical note, model context and service information rather than product photography.

Page summary
Key service context for quote intake and technical review
  • Published in the Rongtao Medical Blog under Repair Case.
  • Relevant system context: GE VIVID ULTRASOUND.
  • Technical focus: Probe / Transducer.
  • Part reference: A00116A1.
  • Use this page to compare product, repair and troubleshooting context before opening a quote request.
BrandGE
Model / systemVIVID ULTRASOUND · VIVID PHILIPS · EPIQ
Part typeProbe / Transducer
Part numberA00116A1
Content typeRepair Case
Technical notes
Technical body content

Fault phenomenon

Philips IU22 Ultrasound crashes during use and cannot be started after a power outage. The operation staff of the department reported that the system crashed during use of IU22 at first, and the system crashed after being shut down and waited for 20 minutes. After a period of use, the system crashed again. The cycle repeats. After a sudden power failure occurs when the system crashes The device cannot be started.

Troubleshooting

The IU22 ultrasound cannot be started. First consider whether the power supply is normal, so first check that the input voltage 220 V, AC, and the stabilized power supply line are normal, and eliminate the abnormal power supply. Because the IU22 ultrasound encounters a short power failure during the crash, it is suspected that the power failure may cause the IU22 to enter the over-voltage protection state, so the capacitor voltage is released after the power failure, and then the power is turned on again for testing. The IU22 resumes and starts but stays The “Loading” interface cannot enter the system. Contact the manufacturer’s engineer to suspect that the system file is lost due to the power failure. It is recommended to reinstall the system, find the local system CD and reinstall the system, and the IU22 will start normally. Considering that the operator has reported repeated crashes, based on previous maintenance experience and the description of the operator, it is suspected that the IU22 ultrasound has poor heat dissipation, causing the phenomenon to reciprocate. Therefore, the casing is removed and it is found that dust has indeed accumulated in the casing. Then, clean the fan, remove the dust, clean the filter, and dry the filter. After installation, the IU22 starts normally, and observes the later use process. No system crashes during use. So far, the fault is repaired.

Service context

Rongtao uses this technical record for GE VIVID ULTRASOUND Probe / Transducer as supporting context for quote intake, photo comparison, board identification and repair-route selection. For live availability, send the machine model, board label, fault symptom and clear photos of both sides of the part.

Tags and related links
Current internal destinations only
Related pages
Same category, brand, model or tag context
Repair case
Repair Cases · Repair case

hitachi aloka f75

GE, VIVID ULTRASOUND, Probe / Transducer, PN EP556600 repair case with part identification, model context and Rongtao quote or repair support.

Reviewed May 18, 2026EP556600
Repair case
Repair Cases · Repair case

hitachi aloka noblusbrief introduction

GE, VIVID ULTRASOUND, Probe / Transducer, PN A00126A1 repair case with part identification, model context and Rongtao quote or repair support.

Reviewed May 18, 2026A00126A1